FDA Warning Letter Criticises Handling of OOS Results at a Contract Laboratory
Recommendation

Tuesday, 10 November 2026 9 .00 - 17.00 h
The U.S. Food and Drug Administration criticises the handling of Out-of-Specification (OOS) results at a U.S.-based contract testing laboratory in a recent Warning Letter. The letter, dated 12 November 2025, was published on the FDA’s website on 02 December 2025 and refers to an inspection conducted in June 2025.
Specifically, the FDA cites the following OOS-related deficiencies:
- Multiple microbiological OOS results were not adequately investigated.
- As a recurring root cause, an “unknown laboratory error” was stated – without sufficient scientific justification.
- The firm relied on retesting without first adequately investigating potential contamination sources.
- The effectiveness of the proposed CAPA was not evaluated.
- The firm lacked a formal, robust OOS management system.
In addition, the agency criticises the overall inadequate quality management. The Quality Unit lacked adequate procedures for deviations, change control and cleaning, and did not ensure effective oversight of laboratory processes. The FDA requests a comprehensive remediation and quality system programme, including an independent retrospective review of all invalidated OOS results for U.S. products over the past three years.
Further information is available in the Warning Letter to Rhyz Analytical Labs published on the FDA website.
Related GMP News
02.11.2026FDA Warning Letter: When Test Procedures Allow Retesting Until an Acceptable Result is Obtained
21.10.2026EMA Updates Information Page on Bioequivalence Guidelines
21.10.2026BP Publishes new Draft Texts
21.10.2026FDA Identifies Critical Deficiencies in Analytical Testing and CGMP Oversight
07.10.2026Analytical Quality Group Developments May through August 2026
07.10.2026Two New FDA Warning Letters Put OOS Investigations in Focus